OilLab 670 Cleveland
Standards
ASTM D92, DIN 51376, EN 22592 (obs.), IP 36ISO 2592
CE Marked
Subject
Flash and Fire Point on petroleum products, gas oils, fuel oils, lubricants. Suitable for flash and fire point detection on different substances and waste materials, having a flash point over 79°C.
Measuring Cleveland Principle
The sample is warmed up according the methods. When the sample reaches the selected test temperature, the flame is passed automatically above the sample. When the flash point is reached, the detection is done by an ionisation detector. For fire point detection, the sample continues to be heated until permanent flame is detected by the second PT100 probe, then the auto extinguisher will be placed on the top of the test cup.
Measuring Cleveland Devices
Analyser equipped with automatic flame exposure device
Measurement of the Flash Point detected by an ionisation detector
Analyser equipped with 2 electrical ignitors and a pilot flame
Measurement of the Fire Point detected by PT100 detector
Measuring Temperature Probe
Platinum resistance PT100 class A
Measuring Parameters
Temperatures: in °C
Measuring range: +79°C … +400°C
Resolution: 0.06 °C
Accuracy: ± 0.1 °C
Repeatability / Reproducibility: as per standards methods or better
Software Features
User friendly interface
All analytical parameters recorded
Customizable analysis parameters and methods
Customizable results report
Printable graphs and results
Self-identification of the typology of the analysers connected
The software includes:
Analysis Menu
Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference
Unknow sample
Audible alarm and displayed messages at the end of the analysis and in case of errors and/or malfunctions
Diagnostic Menu
Direct access to all analog, digital, inputs and outputs
Selectable value displaying: °C / Volt
Calibration Menu
Automatic calibration of each temperature probe
Last calibration date referred to each single probe displayed and relative data printable
Display of calibration diagram
Insertion of offset values
Standard and advanced calibration modes
Data Utilities
Fields for introduction of operator and product name
Archive viewer for files recall
All analysis stored in Excel® compatible format
Storage capacity for more than 60’000 analysis
LIMS compatible
Integrated Touch Screen Panel PC
TFT/LCD 8"
Resolution 1024 × 768, 16.2 M colours
2 USB ports for connection to an external printer and/or external PC
Storage capacity for more than 60'000 analysis
Test Cup
The cup is made of chromium plated brass provided with high temperature resistant handle
Sample level mark
The test cup closing system is totally automatic and grant the maximal security as well as the two electrical ignitors which grant the continuous presence of the test flame.
The Flash Point detection system, which is composed by a ring sensor for the ionization’s determination, constitutes the essential component granting high repeatability and excellent results. Furthermore, a sensor detects the presence of the flame for fire point determination. The device are mounted on a mechanical arm with automatic positioning during analysis.
The test flame pilot is equipped with a flame size regulator as well as a position and direction device for a fine adjustment.
The head can also be equipped with a paddle used to move the surface and perform tests on bitumen.
Heating
Electrical heater
Equipped with over temperature cut-out
Electrical Supply
220V ± 15% / 50 to 60 Hz
115V ± 15% / 60 Hz
Cord Cable
3 conductors flexible cable with schuko plug
Ambient Temperature
Max 35°C
H.R. 80%
Dimensions
width 48 cm
depth 37 cm
height 61 cm
Weight
32 kg
Accessories
LAB-670-12-03: tools kit for bitumen made up of electric drive, support for the PTFE blade flexible transmission with joint and switch.
Spare Parts
LAB-670/05-13: heater (heating plate)
LAB-670/05-26: PT100 for fire point detection
LAB-670/06-21: gas valve
LAB-670/07-01: electrical ignitor, pack of 2 pcs
LAB-670/07-02: gas ignitor
LAB-670/07-03: micro switch
LAB-670/07-04: handle
LAB-670/08-12: PT100 for flash point detection
LAB-670/08-13: detection / ionisation cable
LAB-670/09-04: gas reducer
LAB-670/09-05: calibrated chromium plated brass crucible
LAB-670/10-04: PCB fuses, box of 10
LAB-670/10-05: main electronic board
LAB-670/11-01: silicon tubing, 1 meter
LAB-670/12-01: voltage transformer for ignitor
Calibration Tools
OilLab 80: calibration decade box – PT100 simulator
OilLab 81: set of connectors and cables for cold range