OilLab 690 Tag Flash Point​
Standards
ASTM D56, D3278, D3934, D3941, IP 304, IP 491, IP 492, ISO 1516, ISO 1523, ISO 3679
CE marked
Subject
Flash point on petroleum products having a flash point between ambient temperature and +93°C. Suitable for flash point detection on different substances and (NO) waste materials, solvents…
Measuring Tag Principle
The sample is warmed up according to the methods. When the sample reaches the selected test temperature, the shutter is opened and the ignition system introduces itself automatically. If the flash point is reached, the detection is done by an ionisation detector. If not, the shutter closes again and the sample continues to warm up until the next test temperature.
Measuring Tag Devices
Measurement of the Flash Point detected by an ionisation detector
Testing unit equipped with two ignition systems
Electrical system or flame exposure device
Built-in barometric sensor with automatic barometric correction of results executed by the software
Measuring Temperature Probe
Platinum resistance PT100 class A
Measuring Parameters
Temperatures: in °C
Measuring range: -50°C … +100°C
Testing range: +9°C … +93°C as per test methods
Resolution: 0.06 °C
Accuracy: ± 0.1 °C
Repeatability / Reproducibility: as per standards methods or better
Software Features
User friendly interface
All analytical parameters recorded
Customizable analysis parameters and methods
Customizable results report
Printable graphs and results
Self-identification of the typology of the analysers connected
The software includes:
Analysis Menu
Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference
Unknow sample
Audible alarm and displayed messages at the end of the analysis and in case of errors and/or malfunctions
Diagnostic Menu
Direct access to all analog, digital, inputs and outputs
Selectable value displaying: °C / Volt
Calibration Menu
Automatic calibration of each temperature probe
Last calibration date referred to each single probe displayed and relative data printable
Display of calibration diagram
Insertion of offset values
Standard and advanced calibration modes
Data Utilities
Fields for introduction of operator and product name
Archive viewer for files recall
All analysis stored in Excel® compatible format
Storage capacity for more than 60’000 analysis
LIMS compatible
Integrated Touch Screen Panel PC
TFT/LCD 8"
Resolution 1024 × 768, 16.2 M colours
USB port for connection to an external printer and/or external PC
Storage capacity for more than 60'000 analysis
Test Cup
The cup is made of chromium plated brass provided with high temperature resistant handle
Sample level mark
Heating
Electrical heater
Equipped with over temperature cut-out
Cooling System
Liquid refrigerant controlled by internal solenoid valve
Ignition System
Gas or electric lighter
Shutter
Automatic mechanism opening the shutter conform to the methods
Electrical Supply
220V ± 15% / 50 to 60 Hz
115V ± 15% / 60 Hz
Cord Cable
3 conductors flexible cable with schuko plug
Ambient Temperature
Max 35°C
H.R. 80%
Dimensions
width 48 cm
depth 37 cm
height 61 cm
Weight
32 kg
Spare Parts
LAB-690/05-13: heater
LAB-690/05-16: PT100 bath
LAB-690/06-21: gas valve
LAB-690/07-01: electrical ignitor
LAB-690/07-03: micro switch
LAB-690/07-04: handle
LAB-690/07-05: gas ignitor
LAB-690/08-12: PT100 product
LAB-690/08-13: detection / ionisation cable
LAB-690/09-04: gas reducer
LAB-690/09-05: calibrated brass crucible
LAB-690/09-06: calibrated brass crucible complete with movement
LAB-690/09-07: cover cup movement only
LAB-690/10-04: PCB fuses, Box of 10
LAB-690/10-05: main electronic board
LAB-690/11-01: silicon tubing, 1 meter
LAB-690/12-01: voltage transformer for ignitor
LAB-690/20-01: support PT100 Teflon
Calibration Tools
OilLab 80: calibration decade box – PT100 simulator
OilLab 81: set of connectors and cables for cold range