NewLab 100 Cloud Point

Standards

ASTM D2500,  ASTM D5771, ASTM D5772, ASTM D5773, DIN 51597,  IP 219, IP 444,  IP 445, IP 446, ISO 3015

 

CE marked

 

Subject

Cloud Point of petroleum products and biodiesel fuels

 

Measuring Cloud Point Principle

The sample is cooled down according to the methods while the clouds appearance is observed on the silver bottom of the test jar by means of an optical sensor. The measurement is done by reflection on the silver bottom of the test jar via a fast light detector. The signal from light detector is traded by the LabLink software. The dynamic measurement is performed regardless of the sample’s colour.

 

Measuring Cloud Point Devices

Light pulsed emission on I.R spectrum through a coaxial fibber optic.

 

Measuring Temperature Probe

Platinum resistance PT100 class A

The PT100 is touching the bottom of the test jar

 

Measuring Parameters 

Temperatures: in °C

Measuring range: +80°C … -80°C

Resolution: 0.06 °C

Accuracy: ± 0.1 °C

Repeatability / Reproducibility: as per standards methods or better

 

Software Features

New LabLink software able to manage up to 6 analytical heads simultaneously (stand alone)

 

User friendly interface

All analytical parameters recorded

Customizable analysis parameters and methods

Customizable results report

Printable graphs and results

Self-identification of the typology of the analysers connected

 

The software includes:

Analysis Menu

Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference

Optional methods:

 

fast bath (to reduce the time of analysis)

T-sample – T-bath (Delta T constant)

cooling rate °C / h

Audible alarm and displayed messages (at the end of the analysis and in case of errors and/or malfunctions)

 

Diagnostic Menu

Direct access to all analog, digital, inputs and outputs

Selectable value displaying: °C / Volt

 

Calibration Menu

Automatic calibration of each temperature probe

Last calibration date referred to each single probe displayed and relative data printable

Display of calibration diagram

Insertion of offset values

Standard and advanced calibration modes

 

Data Utilities

Fields for introduction of operator and product name

Archive viewer for files recall

All analysis stored in Excel® compatible format

Storage capacity for more than 60’000 analysis

LIMS compatible

 

Integrated Touch Screen Panel PC

TFT/LCD 12"

Resolution 1024 × 768, 16.2 M colours

2 USB ports for connection to an external printer and/or external PC

Storage capacity for more than 60'000 analysis

 

Test Jar

Same dimensions and volume as described by the standard test methods

Product level mark

Small edge on the top in order to fix the glass cell to the analytical head

Silvered bottom with anti-scratch film protection

 

Cooling System

Integrated gas CFC free motor compressors:

Single stage (for temperatures up to -40°C / 1)

Double stage (for temperatures up to -80°C / 2)

Equipped with an automatic energy power save system. After 15 minutes from the end of the analysis the cooling system goes in stand-by mode.

 

Safety Devices

Pressure controller for 1st stage motor compressor

Pressure controller for 2nd stage motor compressor

Thermostat for 2nd stage activation

Thermo-switch for each cooling / heating jacket

Motor compressors equipped with internal overload devices

 

Electrical Supply

220V ± 15% / 50 to 60 Hz

115V ± 15% / 60 Hz

 

Cord Cable

3 conductors flexible cable 2 m (7 feet) length with PVC sheath oil and heat resistant as per CENELEC directives

 

Ambient Temperature

Max 32 °C

H.R. 80%