NewLab 300 Pour Point

Standards

ASTM D97, ASTM D5853, ASTM D5950, IP 15, IP 441, ISO 3016

 

CE marked

 

Subject

Pour Point of petroleum products, crude oils, motor and engine oils, additives, lubricating oils, …

 

Measuring Pour Point Principle

According to the methods, the sample is cooled down at a specified rate and, at the prescribed temperature intervals, the mechanical arm of the analyser lifts the test jar from the cooling jacket and tilts it in order to bring it in horizontal position to test the flow of the product. The sample movement is detected by the thermal probes (PT100 detection) placed above the sample surface which react if touched by the cooled sample.

 

Measuring Pour Point Devices

  • Two PT100 detection probes placed on the surface of the product

  • Mechanical moving arm bringing the test jar in horizontal position

 

Measuring Temperature Probe

  • Platinum resistance PT100 class A

 

Measuring Parameters

  • Temperatures: in °C

  • Measuring range: +80°C … -80°C

  • Resolution: 0.06 °C

  • Accuracy: ± 0.1 °C

  • Repeatability / Reproducibility: as per standards methods or better

 

Software Features

  • New LabLink software able to manage up to 6 analytical heads simultaneously (stand alone)

 

User friendly interface

  • All analytical parameters recorded

  • Customizable analysis parameters and methods

  • Customizable results report

  • Printable graphs and results

  • Self-identification of the typology of the analysers connected

 

The software includes:

Analysis Menu

  • Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference:

  • (internal) with sample pre-heating

  • (external) without sample pre-heating

 

Optional methods:

  • T-sample – T-bath (Delta T constant)

  • cooling rate °C / h

  • selectable bath steps

  • fast bath

  • selectable tilt out test temperature

  • Audible alarm and displayed messages (at the end of the analysis and in case of errors and/or malfunctions)

 

Diagnostic Menu

  • Direct access to all analog, digital, inputs and outputs

  • Selectable value displaying: °C / Volt

 

Calibration Menu

  • Automatic calibration of each temperature probe

  • Last calibration date referred to each single probe displayed and relative data printable

  • Display of calibration diagram

  • Insertion of offset values

  • Standard and advanced calibration modes

 

Data Utilities

  • Fields for introduction of operator and product name

  • Archive viewer for files recall

  • All analysis stored in Excel® compatible format

  • Storage capacity for more than 60’000 analysis

  • LIMS compatible

 

Integrated Touch Screen Panel PC

  • TFT/LCD 12"

  • Resolution 1024 × 768, 16.2 M colours

  • 2 USB ports for connection to an external printer and/or external PC

  • Storage capacity for more than 60'000 analysis

 

Test Jar

  • Same dimensions and volume as described by the standard test methods

  • Product level mark

  • Small edge on the top in order to fix the glass cell to the analytical head

 

Cooling System

  • Integrated gas CFC free motor compressors:

  • Single stage (for temperatures up to -40°C / 1)

  • Double stage (for temperatures up to -80°C / 2)

  • Equipped with an automatic energy power save system. After 15 minutes from the end of the analysis the cooling system goes in stand-by mode.

 

Safety Devices

  • Pressure controller for 1st stage motor compressor

  • Pressure controller for 2nd stage motor compressor

  • Thermostat for 2nd stage activation

  • Thermo-switch for each cooling / heating jacket

  • Motor compressors equipped with internal overload devices

 

Electrical Supply

  • 220V ± 15% / 50 to 60 Hz

  • 115V ± 15% / 60 Hz

 

Cord Cable

  • 3 conductors flexible cable 2 m (7 feet) length with PVC sheath oil and heat resistant as per CENELEC directives

 

Ambient Temperature

  • Max 32 °C

  • H.R. 80%

 

Spare Parts

LAB-xxx/005-03: heather + auto adhesive + insulation

LAB-xxx/005-04: thermo switch

LAB-xxx/005-06: PT100 bath

LAB-xxx/007-02: static relay

LAB-xxx/007-04: PCB fuse 1 AT (box of 10 pcs)

LAB-xxx/006-01: cooling fluid valve + fitting

LAB-300/007-01:               main electronic board Pour Point

LAB-300/002-16:               precision potentiometer

LAB-300/008-12:               PT100 product w/connector

LAB-300/008-13:               PT100 detection

LAB-300/008-04:               calibrated test jar

LAB-300/008-041: o-ring for test jar

 

Calibration Tools

OilLab 80: calibration decade box – PT100 simulator

OilLab 81: set of connectors and cables for cold range

Downloads

 

Newlab 300

 

Alternative Products

 

A sentence here about related products with image links below to products you want customer to consider. 

SUBMIT

Name:

This field is required.

Thank You for your enquiry.

Your information has been sent successfully.  One of the Crea Laboratory technologies Team will be in contact with your shortly. 

 

If your enquiry is urgent, please call us on 1300 552 725

Email:

This field is required.

Phone:

This field is required.

Industry:

This field is required.

Company Name:

This field is required.

Address:

This field is required.

Post Code:

Country:

This fi.

NEWLAB 300

For more information on this product or to arrange a call from one of the Cea Petrochemical Team, please complete the contact form below.