NewLab 410 Freezing Point

Standards

ASTM D852, ASTM D1177, ASTM D1493, ASTM D1655, ASTM D2386, ASTM D5901, ASTM D5972, ASTM D7154, IP 16, IP 435, IP 528, IP 529, ISO 3013

 

CE marked

 

Subject

Freezing Point of aviation fuels, aviation gasoline, aviation turbine fuels, engine coolants, antifreeze products, brake fluids, …

Solidification Point of Benzene.

Solidification Point of Industrial Organic Chemicals.

 

Measuring Freezing Point Principle

According to the methods, the sample is cooled down and stirred. The solid hydrocarbon crystals formation are detected by means of a light beam throught fiber optic reflected thanks to a mirror. As soon as crystals are detected, the sample is warmed up until their complete disappearance.

 

Measuring Freezing Point Devices

Light pulsed emission on I.R spectrum through a coaxial fibber optic

Coaxial fibber optic equipped with a mirror

 

Measuring Temperature Probe

Platinum resistance PT100 class A

 

Stirrer

A micro-motor drives all the mechanical system

3 coils stirrer made of brass

 

Measuring Parameters

Temperatures: in °C / °F

Measuring range: +80°C … -100°C

Resolution: 0.06 °C

Accuracy: ± 0.1 °C

Repeatability / Reproducibility: as per standards methods or better

 

Software Features

New LabLink software able to manage up to 6 analytical heads simultaneously (stand alone)

User friendly interface

All analytical parameters recorded

Customizable analysis parameters and methods

Results report

Printable graphs and results - any Windows® compatible printer can be used

 

The software includes:

Analysis Menu

Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference

 

Optional methods:

  • special detection of contaminants

 

Audible alarm and displayed messages at the end of the analysis and in case of errors and/or malfunctions

 

Diagnostic Menu

  • Direct access to all analog, digital, inputs and outputs

  • Selectable value displaying: °C / °F / Volt

 

Calibration Menu

  • Automatic calibration of each temperature probe

  • Last calibration date referred to each single probe displayed and relative data printable

  • Display of calibration diagram

  • Insertion of offset values

  • Standard and advanced calibration modes up to 100 calibration points

 

Data Utilities

  • Fields for introduction of operator and product name

  • Archive viewer for files recall

  • All analysis stored in Excel® compatible format and JPG image

  • Storage capacity for more than 60’000 analysis

  • LIMS compatible

 

Integrated Touch Screen Panel PC

TFT/LCD 12"

Resolution 1024 × 768, 16.2 M colours

2 USB ports for connection to an external printer and/or external PC

Storage capacity for more than 60'000 analysis

 

Test Jar

Product level

  • Same dimensions and volume as described by the standard test methods

  • mark at 25 ml

  • Small edge on the top in order to fix the glass cell to the analytical head

 

Cooling System

Insulated cooling jackets

Integrated gas CFC free motor compressors:

  • Double stage (for temperatures up to -80°C / 2)

  • Equipped with an automatic energy power save system. After 15 minutes from the end of the analysis the cooling system goes in stand-by mode.

 

Safety Devices

Pressure

  • controller for 1st stage motor compressorPressure controller for 2nd stage motor compressor

  • Thermostat for 2nd stage activation

  • Thermo-switch for each cooling / heating jacket

  • Motor compressors equipped with internal overload devices

 

Electrical Supply

  • 220V ± 15% / 50 to 60 Hz

  • 115V ± 15% / 60 Hz

 

Cord Cable

  • 3 conductors flexible cable 2 m (7 feet) length with PVC sheath oil and heat

 

Ambient Temperature

  • Max 32 °C

  • H.R. 80%

 

Spare Parts

LAB-xxx/005-03: heather + auto adhesive + insulation

LAB-xxx/005-04: thermo switch

LAB-xxx/005-06: PT100 bath

LAB-xxx/007-02: static relay

LAB-xxx/007-04: PCB fuse 1 AT (box of 10 pcs)

LAB-xxx/006-01: cooling fluid valve + fitting

LAB-400/007-01: main electronic board Freezing Point

LAB-400/008-04: PT100 product w/connector

LAB-400/008-05: stirrer

LAB-400/008-08: mirror for Freezing Point

LAB-400/008-06: motor for stirrer

LAB-400/008-07: fibber optic for Freezing Point

LAB-400/008-09: electronic board for detection

LAB-410/008-12: removable glass cell for Freezing Point

LAB-410/008-041: o-ring for Freezing Point test jar

 

Calibration Tools

OilLab 80: calibration decade box – PT100 simulator

OilLab 81: set of connectors and cables for cold range

Downloads

 

Newlab 410

 

Alternative Products

 

A sentence here about related products with image links below to products you want customer to consider. 

SUBMIT

Name:

This field is required.

Thank You for your enquiry.

Your information has been sent successfully.  One of the Crea Laboratory technologies Team will be in contact with your shortly. 

 

If your enquiry is urgent, please call us on 1300 552 725

Email:

This field is required.

Phone:

This field is required.

Industry:

This field is required.

Company Name:

This field is required.

Address:

This field is required.

Post Code:

Country:

This fi.

NEWLAB 410

For more information on this product or to arrange a call from one of the Cea Petrochemical Team, please complete the contact form below.