OilLab 620 RECC Rapid Equilibrium Closed Cup​
Standards
ASTM D3828, IP 303, EN ISONF3679
Marchio CE
Subject
These test methods cover procedures for the determination of the flash point by a small scale closed tester. The procedures may be used to determine the actual flash point temperature of a sample or whether a product will or will not flash at a specified temperature (flash/no flash).
Measuring R.E.C.C. Principle
A specimen of a sample is introduced by means of a syringe into the cup of the selected apparatus that is set and maintained at the specific temperature/expected flash point. After a specified time a test flame is applied and the observation made whether or not a flash occurred.
Measuring R.E.C.C. Devices
Testing unit equipped with two ignition systems
Electrical system or flame exposure device
Flash point detected by a ionization system
Measuring Temperature Probe
Platinum resistance PT100 class A
Measuring Parameters
- Temperatures: in °C
- Measuring range: –50°C … +350°C
- Resolution: 0.06 °C
- Accuracy: ± 0.1 °C
- Repeatability / Reproducibility: as per standards methods or better
LabLink operating Software
User friendly interface
All analytical parameters recorded
Customizable analysis parameters and methods
Customizable results report
Printable graphs and results
Self-identification of the typology of the analysers connected
The software includes:
Analysis Menu
Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference
Unknow sample
Audible alarm and displayed messages at the end of the analysis and in case of errors and/or malfunctions
Diagnostic Menu
Direct access to all analog, digital, inputs and outputs
Selectable value displaying: °C / Volt
Calibration Menu
Automatic calibration of each temperature probe
Last calibration date referred to each single probe displayed and relative data printable
Display of calibration diagram
Insertion of offset values
Standard and advanced calibration modes
Data Utilities
Fields for introduction of operator and product name
Archive viewer for files recall
All analysis stored in Excel® compatible format
Storage capacity for more than 60’000 analysis
LIMS compatible
Integrated Touch Screen Panel PC
- TFT/LCD 8"
- Resolution 1024 × 768, 16.2 M colours
- USB port for connection to an external printer and/or external PC
- Storage capacity for more than 60'000 analysis
Test Cup
The cup is made of aluminium and equipped with high temperature resistant o-ring.
Automatic opening, closing and positioning of the sliding shutter.
The test cup is completely made of aluminium and is equipped with a high temperatures resistant o-ring, allowing an uniform sealing of the closing cover.
Particular attention has been paid to the heating system allowing the best heat’s transmission without dispersion on air. The power of the heater is therefore of only 420W but allows to obtain a temperature higher than 370°C.Particular attention has been paid to the heating system allowing the best heat’s transmission without dispersion on air. The power of the heater is therefore of only 420W but allows to obtain a temperature higher than 370°C.
The ionisation components are the essential flash point detection system that grants excellent results and high repeatability. The quantity of sample (2 ml / 4 ml) is injected into the cup throught the filling orifice.The instrument is equipped with two ignition systems. An electric pilot that ignites the test flame 30 seconds before the test.
Heating
- Electrical heating cartridges
- Equipped with over temperature cut-outCooling SystemAir fan
Warning light and acoustic signal
When the test temperature is reached, the light blink and an acoustic signal is emitted to inform the operator that the sample must be injected.If the injection of the sample is not performed and confirmed the signal will be repeated after 30 seconds.
Shutter
- Automatic mechanism opening the shutter conform to the methods
Electrical Supply
- 220V ± 15% / 50 to 60 Hz
- 115V ± 15% / 60 Hz
Cord Cable
- 3 conductors flexible cable with
schuko plug
Ambient Temperature
- Max 35°C
- H.R. 80%
Dimensions
- width 31 cm
- depth 47 cm
- height 52 cm
Weight
- 27 kg
Spare Parts
LAB-620/05-13: heating cartridges
LAB-620/06-21: gas valve
LAB-620/07-01: electrical ignitor
LAB-620/07-03: micro switch
LAB-620/07-05: gas ignitor
LAB-620/08-12: PT100 sensor
LAB-620/08-13: detection / Ionisation cable
LAB-620/09-04: gas reducer
LAB-620/10-04: PCB fuses, box of 10
LAB-620/10-05: main electronic board
LAB-620/11-01: silicon tubing, 1 meter
LAB-620/12-01: voltage transformer for ignitor
LAB-620/13-01: high temperature resistant o-ring for cup
Calibration Tools
OilLab 80: calibration decade box – PT100 simulator
OilLab 81: set of connectors and cables for cold range